Indian Journal of Clinical and Experimental Ophthalmology

Print ISSN: 2395-1443

Online ISSN: 2395-1451

CODEN : IJCEKF

Indian Journal of Clinical and Experimental Ophthalmology (IJCEO) is open access, a peer-reviewed medical journal, published quarterly, online, and in print, by the Innovative Education and Scientific Research Foundation (IESRF) since 2015. To fulfil our aim of rapid dissemination of knowledge, we publish articles ‘Ahead of Print’ on acceptance. In addition, the journal allows free access (Open Access) to its content, which is likely to attract more readers and citations of articles published in IJCEO. Manuscripts must be prepared in more...

Article type

Original Article


Article page

209- 213


Authors Details

Sunny Biswas*, Prakash Kumar, Md. Danish Siddiqui, Shorubha Dinakaran, Shankar Das


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 649

PDF Downloaded: 354


A brief history of the evolution of tonometers and comparison of goldmann applanation tonometer versus non–contact tonometer and their correlation to central corneal thickness


Original Article

Author Details : Sunny Biswas*, Prakash Kumar, Md. Danish Siddiqui, Shorubha Dinakaran, Shankar Das

Volume : 9, Issue : 2, Year : 2023

Article Page : 209-213

https://doi.org/10.18231/j.ijceo.2023.040



Suggest article by email

Get Permission

Abstract

Purpose: This study was designed to evaluate and compare differences between Goldmann applanation tonometer’s (GAT) values and non-contact tonometer’s (NCT) values In various central corneal thickness(CCT).
Materials and Methods : This prospective study of 100 eyes of 50 subjects was done for the age group of 25 yrs – 75 yrs, who did not have any glaucomatous changes in the optic disc.
Intraocular pressure readings were taken by GAT and NCT was measured within a fixed time range of 10:00 AM to 1:00 PM to minimize the effect of diurnal variation. CCT and NCT readings were measured by NIDEK 530P while GAT was performed by Appasamy ref AATM-K001 after 20 minutes and was recorded from both eyes of 50 subjects.
Then the CCT effect was correlated with the intraocular pressure values, which were obtained by GAT and NCT.
Result: The data were recorded from 100 eyes of 50 subjects, of whom 24 were females and 26 were males.
The mean age was 51.22 ± 2.7 years, while the mean CCT value was 530.6 ± 5.7 µ.
The mean IOP values respectively for NCT and GAT were 15.41 ± 0.6 mmHg, and 15.11 ± 0.5 mmHg, respectively (Table 1).
The data shows a significantly strong positive correlation between NCT and CCT (r= 0.73, P< .0001), and a strong positive correlation between GAT and CCT (r= 0.55, P< .0001) (Table 2, 5 and 6).
The Pearson coefficient value between NCT and GAT was also strong (r = 0.76, P< .00001)(Table 3).
Conclusion: The findings in our study correlate with earlier research. It demonstrates that, with the application of appropriate correction factors for CCT, NCT can be considered a good screening tool for the measuring Intraocular pressure.
 

Keywords: Goldman Applanation tonometer (GAT), Non­Contact Tonometer (NCT), Central Corneal Thickness (CCT), Tonometer, Intraocular Pressure (IOP).


How to cite : Biswas S, Kumar P, Siddiqui M D, Dinakaran S, Das S, A brief history of the evolution of tonometers and comparison of goldmann applanation tonometer versus non–contact tonometer and their correlation to central corneal thickness. Indian J Clin Exp Ophthalmol 2023;9(2):209-213

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.