Indian Journal of Clinical and Experimental Ophthalmology

Print ISSN: 2395-1443

Online ISSN: 2395-1451

CODEN : IJCEKF

Indian Journal of Clinical and Experimental Ophthalmology (IJCEO) is open access, a peer-reviewed medical journal, published quarterly, online, and in print, by the Innovative Education and Scientific Research Foundation (IESRF) since 2015. To fulfil our aim of rapid dissemination of knowledge, we publish articles ‘Ahead of Print’ on acceptance. In addition, the journal allows free access (Open Access) to its content, which is likely to attract more readers and citations of articles published in IJCEO. Manuscripts must be prepared in more...

Article type

Original Article


Article page

492-498


Authors Details

Aditi Sharma*, Aditya Vikram Sharma


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1412

PDF Downloaded: 865


Comparison of accuracy between ultrasound B scan and partial coherence interferometry (IOL master) in IOL (intraocular lens) power calculation


Original Article

Author Details : Aditi Sharma*, Aditya Vikram Sharma

Volume : 4, Issue : 4, Year : 2018

Article Page : 492-498

https://doi.org/10.18231/2395-1451.2018.0110



Suggest article by email

Get Permission

Abstract

Aims and Objectives: To compare the accuracy of ultrasound A scan biometry and IOL master (PCI) biometry in intraocular power calculation, by doing a comparative study of difference in biometry by conventional methods and optical biometry (IOL master).
Materials and Methods: Place of study: Aravind Eye Hospital, India.
Design of Study: Prospective, non randomised, clinical study.
Study Group and Duration: The study involves patients who underwent cataract surgeries at Aravind Eye Hospital, Tirunelveli during the specified time period.
The IOL power and predicted refractive outcome were calculated for every patient by both the methods, A scan ultrasound Biometry IOL master (Partial Coherence interferometry)
Inclusion Criteria: Only those patients in whom it was possible to secure a reliable IOL master reading (on the basis of good SNR), were included.
Results: In a 12 month period, 337 eyes were consecutively enrolled in the study. Both ultrasound A scan and partial coherence interferometry was performed and IOL power implanted as per the IOL master.

Inclusion criterion and exclusion criterion used as mentioned.
The difference in between the final and predicted refraction is compared for both the groups. The overall Mean absolute error was 0.402 ± 0.407 with the ultrasound A scan & 0.388 ± 0.406 for the IOL master, with the probability value (p-value) 0.3802.
The mean numerical error (the difference between the refractive outcome and the predicted spherical equivalent) is -0.226 ± 0.526 (min -2.66 & max 2.06) for the ultrasound scan group and -0.217 ± 0.518 (min -2.41 & max 2.10) for the IOL master group, with the probability value (p-value) being 0.514.
We got a comparative result in our study, though not significant, but it is more in favour of IOL master (PCI) as per the mean absolute error and mean numerical error calculations.
When analysing the Mean Axial Length readings, the findings were almost similar, with an insignificant p-value.

Keywords: Biometry, IOL master, Power calculation, Ultrasound.


How to cite : Sharma A, Sharma A V, Comparison of accuracy between ultrasound B scan and partial coherence interferometry (IOL master) in IOL (intraocular lens) power calculation. Indian J Clin Exp Ophthalmol 2018;4(4):492-498

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.