Indian Journal of Clinical and Experimental Ophthalmology (IJCEO) is open access, a peer-reviewed medical journal, published quarterly, online, and in print, by the Innovative Education and Scientific Research Foundation (IESRF) since 2015. To fulfil our aim of rapid dissemination of knowledge, we publish articles ‘Ahead of Print’ on acceptance. In addition, the journal allows free access (Open Access) to its content, which is likely to attract more readers and citations of articles published in IJCEO. Manuscripts must be prepared in more...

Article type

Original Article


Article page

329- 332


Authors Details

CN Madhusudhana, S. Shyam


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1498

PDF Downloaded: 555


Comparison of rebound tonometry with applanation tonometry for intraocular pressure measurement


Original Article

Author Details : CN Madhusudhana, S. Shyam

Volume : 3, Issue : 3, Year : 2017

Article Page : 329-332


Suggest article by email

Get Permission

Abstract

Purpose: To compare the readings of the Goldmann’s applanation tonometry with that of the iCare Ta01i in Indian patients.
Materials and Methods: Intraocular pressure measurements were obtained in 440 eyes of 220 subjects in which 320 were normal eyes and 120 eyes had a high IOP. A thorough examination was done and IOP was measured in all subjects using iCare Ta01 tonometer and the Goldmann’s applanation tonometer 3 times each and the average was taken. All examinations and measurements were done by the same 2 personnel.
Results: Intraocular pressure measurements with ICare Ta01 tonometer compared with the Goldmann’s applanation tonometer were not found to have a significant difference (mean ±2.34mmHg). The 95% confidence interval between readings with the 2 methods was within acceptable limits.
Conclusion: The ICare Ta01 tonometer measurement of intraocular pressure showed excellent correlation with those from the Goldmann applanation tonometer, which is the gold standard for IOP measurement. The iCare, with all its advantages over the GAT, can be used as a routine instrument for measurement of IOP in ophthalmology practice.


How to cite : Madhusudhana C, Shyam S, Comparison of rebound tonometry with applanation tonometry for intraocular pressure measurement. Indian J Clin Exp Ophthalmol 2017;3(3):329-332

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.