Indian Journal of Clinical and Experimental Ophthalmology

Print ISSN: 2395-1443

Online ISSN: 2395-1451

CODEN : IJCEKF

Indian Journal of Clinical and Experimental Ophthalmology (IJCEO) is open access, a peer-reviewed medical journal, published quarterly, online, and in print, by the Innovative Education and Scientific Research Foundation (IESRF) since 2015. To fulfil our aim of rapid dissemination of knowledge, we publish articles ‘Ahead of Print’ on acceptance. In addition, the journal allows free access (Open Access) to its content, which is likely to attract more readers and citations of articles published in IJCEO. Manuscripts must be prepared in more...

Article type

Original Article


Article page

85-88


Authors Details

Hemant Sharma*, Sanjeev K Nainiwal, Akshay Sarraf, Rakesh Porwal, Vijaya Sharma


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 670

PDF Downloaded: 867


Comparative study between schiotz tonometer and goldmann applanation tonometer in glaucomatous and normal individuals


Original Article

Author Details : Hemant Sharma*, Sanjeev K Nainiwal, Akshay Sarraf, Rakesh Porwal, Vijaya Sharma

Volume : 8, Issue : 1, Year : 2022

Article Page : 85-88

https://doi.org/10.18231/j.ijceo.2022.016



Suggest article by email

Get Permission

Abstract

Purpose: Checking of intraocular pressure (IOP) is one of the basic investigations in a general ophthalmology work up. In this study we attempted to compare the IOP readings obtained by Goldmann applanation tonometer and Schiotz indentation tonometer in glaucomatous and normal individuals attending general ophthalmology OPD in a tertiary care centre in Central Rajasthan.
Materials and Methods: It is a case series study in which IOP was checked in patients using two tonometers. Statistical analysis was done to evaluate the agreement between instruments.
Result: In this study, when measured with Goldmann applanation tonometer (GAT), it was found that the mean IOP of both right and left eye in group A (Glaucomatous subjects) (24.19±12.94 and 23.34±13.73 respectively) was significantly higher than mean IOP of both right and left eye in group B(non glaucomatous subjects) (15.06±2.49 and 15.12±2.38 respectively). Likewise, when measured with Shiotz tonometer (ST), mean IOP of both right and left eye in group A (26.05±11.93 and 24.65±12.28 respectively) were very significantly higher in comparison to group B (16.79±2.64 and 16.85±2.47 respectively).
Conclusion: The mean pressures obtained by the Schiotz tonometer were higher than the mean Goldmann pressures which indicate that the Schiotz tonometer tends to read higher than the Goldmann tonometer.
 

Keywords: Glaucoma, Tonometry, Schiotz tonometer, Goldmann applanation tonometer (GAT).


How to cite : Sharma H, Nainiwal S K, Sarraf A, Porwal R, Sharma V, Comparative study between schiotz tonometer and goldmann applanation tonometer in glaucomatous and normal individuals. Indian J Clin Exp Ophthalmol 2022;8(1):85-88

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.