Indian Journal of Clinical and Experimental Ophthalmology

Print ISSN: 2395-1443

Online ISSN: 2395-1451

CODEN : IJCEKF

Indian Journal of Clinical and Experimental Ophthalmology (IJCEO) is open access, a peer-reviewed medical journal, published quarterly, online, and in print, by the Innovative Education and Scientific Research Foundation (IESRF) since 2015. To fulfil our aim of rapid dissemination of knowledge, we publish articles ‘Ahead of Print’ on acceptance. In addition, the journal allows free access (Open Access) to its content, which is likely to attract more readers and citations of articles published in IJCEO. Manuscripts must be prepared in more...

Article type

Original Article


Article page

541-547


Authors Details

Reshma Ramakrishnan, Mamta Agrawal*, Suvarna Kalapad, Saurabh Shrivastava, Priyanka Gandhi


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 454

PDF Downloaded: 487


A study to compare intraocular pressure using Perkins and Schiotz tonometer with respect to Goldmann applanation tonometer as a gold standard in a tertiary health care centre, Navi Mumbai


Original Article

Author Details : Reshma Ramakrishnan, Mamta Agrawal*, Suvarna Kalapad, Saurabh Shrivastava, Priyanka Gandhi

Volume : 7, Issue : 3, Year : 2021

Article Page : 541-547

https://doi.org/10.18231/j.ijceo.2021.107



Suggest article by email

Get Permission

Abstract

Aim: To compare IOP measurement of Perkin’s tonometer and Schiotz tonometer with respect to Goldmann applanation tonometer (GAT).
Materials and Methods: 100 eyes of 50 patients aged 18 to 65 years presenting to Ophthalmology OPD were selected by randomization and detailed general and ophthalmic examination was done. IOP was measured using applanation tonometers (Perkin’s and GAT) and indentation tonometer (Schiotz tonometer) in that order.
Results: The study population included 38 male eyes and 62 female eyes; 50 right eyes and 50 left eyes. There was a statistically highly significant difference seen for the values between the groups (p<0> There was a statistically non significant difference seen for the values between Perkin’s tonometer and GAT (p>0.05). While there was a statistically highly significant difference seen for the values between GAT and Schiotz tonometer (p<0> There was a statistically non significant difference seen for the IOP values between right eye and left eye. There was a statistically non significant difference seen for the IOP values between males and females.
Conclusion: Measurement of IOP with Perkin’s was closer to the values obtained by GAT. Perkin’s being portable, easy to use and precise tonometer than Schiotz tonometer, can be considered as an excellent substitute to GAT, for large scale examination, if cost is not a concern.
 

Keywords: Intraocular pressure, Perkin's tonometer, GAT, Schiotz tonometer.


How to cite : Ramakrishnan R, Agrawal M, Kalapad S, Shrivastava S, Gandhi P, A study to compare intraocular pressure using Perkins and Schiotz tonometer with respect to Goldmann applanation tonometer as a gold standard in a tertiary health care centre, Navi Mumbai. Indian J Clin Exp Ophthalmol 2021;7(3):541-547

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.